Art nr 108491-BS ZyNthAir

Gas Mixer for PAC CID 510 Analyzers

ACCURATE, EASY AND RELIABLE COMBUSTION AIR

The PAC ZyNthAir Gas Mixer is an accessory for existing CID Cetane Analyser that removes the need for synthetic air while producing constant, reliable and accurate combustion air.

The ZynthAir is backward compatible with existing CID users, and many of them will benefit from the reduced cost of purchasing synthetic air and the need to perform a Cetane calibration between gas cylinder swaps. The Gas Mixer requires three gas inputs: Nitrogen and Zero Air used to synthesise the 20% Oxygen for reliable combustion, and then small amounts of Oxygen for the automatic routine calibration of the Oxygen Sensor.

FEATURES AND BENEFITS
- Removes the need for Synthetic Air
- No more Cetane calibration on Cylinder changes
- Plug and play with existing CID systems, no requirement for a software update
- Oxygen content 20+/- 0.05% for high accurate Cetane measurement
- High-performance temperature controlled paramagnetic oxygen sensor with ±0.01% accuracy (intrinsic error)
- Auto-calibration of Oxygen Sensors
- Annual maintenance and calibration of pressure sensors with the last 6 records stored
- Self-diagnostics
- LCD interface shows current status, oxygen level and errors
- Visual LED for indication of status of the Gas Mixer
- Control App allowing for easy set-up maintenance and system review

Download PDF
Physical specifications

Width: 203 mm
Depth: 508 mm
Height: 610 mm
Weight: 34 kg

Power requirements

Input power voltage: 90 to 240 VAC, Indoor use only
Frequency: 50 to 60 Hz
Power Consumption: < 60 W

Gas requirements

Air Supply
- Compressed clean dry air (Zero-Air recommended) Oxygen concentration: 20 - 22 %
- CO2: < 1 ppm
- CO: < 1 ppm
- THC: < 0.1 ppm
- Water: < 2 ppm
- Pressure: 37 – 40 barg

Nitrogen Supply - Compressed Nitrogen
- Purity: ≥ 99.999%
- Pressure: 37 – 40 barg

Oxygen Supply - Compressed Oxygen
- Purity: ≥ 99.994%
- Pressure: 2.2 – 3.4 barg

Output Gas

Combustion Air
- Compressed combustion air
- Oxygen concentration: 20.0% ± 0.05%,
- Pressure: ≤ 25 barg

Environmental conditions & certification

Operating Temperature: 10 to 35°C (50 to 95°F), recommended 15 to 25°C (59 to 77°F)
Operating Humidity: ≤ 85 % - non-condensing
Storage Temperature: -20 to 50°C (-4 to 122°F)
Storage Humidity: ≤ 85 % - non-condensing
Pollution: Pollution Degree 2
Installation: Installation Category II
Altitude: ≤ 2000 m
Ingress Protection: IP 54/NEMA 3
Certification: ISO 9001:2015, CE, ROHS II
There are no accessories available for this product.