Cloud Point Analyzer (ASTM D2500, D5771)
Combining a revolutionary patented built-in cooling system and highly precisedetection mechanism, the ISL OptiCPP performs pour and/cloud point testing of any petroleum product, down to -95C (-139F), in accordance with all the international standards. Its unique cooling technology eliminates external liquid connections, heat and noise but also the maintenance associated with external cooling.100% configurable, OptiCPP accommodates your unique testing requirements, yet also provides strict compliance to international standard test methods. Automation enables unattended operation with excellent reproducibility and andrepeatability.The OptiCPP is ready to work whenever and wherever you are; simply plug in the electrical cord, and begin testing. The unit can be powered from car battery with an optional adapter. Pour point is detected by tilting the sample no turning, pressing or twisting as an optical surface. The detection system precisely monitors the movement of the specimen surface whereas optics detect cloudpoint in the specimen when wax crystals first appear.
FEATURES & BENEFITS
Precise and reliable analysis performance
Ergonomic cordless cloud point and pour point heads
Powerful and flexible operation with ultra-low temperature capabilities
Enhanced calibration features for full compliance and quality assurance
Multiple options to calibrate temperature probes
ASTM D5771
IP 444
In correlation to:
ASTM D2500
ISO 3015
IP 219
JIS K2269
Detection:
Cloud Point: optical
Cooling:
Internal cooling system; stepped cooling according to method or user defined; up to 20 steps can be customized
Test Interval:
Cloud Point: automatic 0.1°C or according to the method
Temperature Range:
Jacket: -105 to +55°C (-157 to +131°F)
Sample: -95 to +51°C * (-139 to 123°F)
samples having EPP higher than +35°C have to be preheated externally
Temperature Measurement:
Jacket: 0.5°C accuracy; Pt100 metal probe sample:
0.1°C accuracy; Pt100 glass (standard) or metal (optional) probe engraved with serial number
Calibration:
Automatic temperature calibration routine 20-point temperature probe correction table Programmable calibration frequency
Utility Requirements:
Electrical: 100–240 VAC, 150W, 50/60 Hz
Dimensions: 30 cm x 65 cm x 34 cm (W x D x H)
Weight: 28 kg
Compatible with IRIS software