Art nr 104000 OptiFlash PM

Automatic PMCC Flash Point Analyser (ASTM D93)

The Herzog OptiFlash Pensky Martens determines the flash point of petroleum products, biodiesels, solvents, fluxed bitumen, food, and beverages up to 400°C. Fire detection in extended range around test cup and flash detection outside the cup.

TEST METHODS:
• ASTMD93 (A,B,C)
• ISO2719 (A,B,C)
• ENISO2719 (A,B,C)
• IP34 (A,B,C)
• JIS K 2265
• GB/T261

ADVANTAGES & FEATURES:
• Built in QC-Chart for trend monitoring
• Sample temperature sensor with built in calibration memory
• User friendly touch screen with alpha numeric key pad
• Flash point measuring range from ambient temperature up to 400°C

OPTIONAL FEATURES:
• Automatic lift arm for installation of the test cup cover with the sensors
• Disconnect cup cover by a simple push button
• Easy cleaning with removable shutter
• High safety standard with ultra-fast fire detection and extinguishing system
• Detection for samples which form Flash outside of the cup


Flash point measuring range: 30°C to 400°C
Heating System: Heater band for uniform heating of the heater block

Temperature Measurement: Intelligent Pt 100 probe with built in calibration (10p)
Thermometer type: Glass or Metal Pt100 probe
Temperature range -50°C to +450°C, resolution 0.1°C
Sample Stirrer: Automatic stirrer, test method or user-defined RPM from 0 to 300
Barometric Pressure Sensor: Built-in sensor for automatic correction of flash point
Flash Detection: Unique thermal flash point detection
User Interface: 7i nch colored touch screen
Result Documentation: 500 results, 200 products, reports for printer and LIMS
Printer: USB interface and PCL5 or higher
LIMS interface: Ethernet or RS232
Data Export: USB memory stick, import into Excel
QC-Functions: Automatic QC-sample handling and QC-chart

Ignition Source:
• Intelligent electric igniter with automatic power management over life time
• Automatic gas ignition with gas monitoring

SAFETY FEATURES
• Fire Extinguisher
• Safety Pre-test
• Adjustable alarm functions
• Password Protection
• User defined calibration intervals and diagnostics

Electrical: 230VAC ±10% with automatic switching; 50-60Hz; 1100 W
Ambient Conditions: Operating temperature: 10°C to 35°C; Humidity: 80% rel. at 35°C
Storage temperature: -15°C to +55°C
Dimensions & Weight: 25 cm wide, 51 cm deep, 56 cm tall, 25kg
1700-000-004 PAC, ISL Barcode reader to enter sample number
634-198 PAC, Herzog Plug 6pol. For Alarm Output
1000-004-001 PAC, Herzog Pt100 probe simulator box 0 to +400°C
104854 PAC, Herzog Upgrade kit gas ignition for PM
104855 PAC, Herzog Upgrade kit gas ignition for PM with Lift Arm
1700-000-001 PAC, Herzog USB cable for printer
1200-004-002 PAC, Herzog USB Printer A4 100V - 230VAC without Cable
102165 PAC, ISL USB Thermal Ticket Printer 100-230V
01000-329-53 PAC, Herzog CRM Flash Point PM Method A app. +210°C, 255ml
01000-329-51 PAC, Herzog CRM Flash Point PM Method A appr. +65°C, 255ml
01000-329-52 PAC, Herzog CRM Flash Point PM Method B app. +100°C, 255ml
103733 PAC, Herzog Electric igniter for OptiFlash
102700-5 PAC, Herzog Set of 5 Thermal Printer Paper Rolls
104852 PAC, Herzog Test Cup PM