Laboratorium instrumenten, Petroleum testapparatuur, Temperatuur, Kalibratie apparatuur, Kalibratie Standaarden, Reactoren & automatisatie, Stockverkoop toestellen, Spare parts At the top of technological performance,
it’s the service that makes all the difference.

labo instruments optimus instruments

As market leader in Petroleum Analyzers, Optimus Instruments provides innovative and cost-effective techniques for physical property testing, distillation and element analysis for almost 30 years. Great instruments combined with excellent after sales service and our worldwide unique ISO17025 accredation for verification makes Optimus Instruments a partner you can count on.

ACM202024EUR - ICOUNT ACM20
AVIATION FUEL CONTAMINATION LASER PARTICLE COUNTER

Delivery To be confirmed.

ACM202024EUR - Aviation fuel contamination laser particle counter

Parker Filtration’s icountACM20 is the benchmark particle counter for use in monitoring the levels of contamination in fuels. With a 2 minute test procedure, the icountACM20 offers multi-standard ISO cleanliness reporting, optimising the useful lifetime of fuel filters so that you get maximum filtration for your money.

Product Features:
icountACM20 monitors aviation fuel contamination to DEFSTAN 91-91 Issue 6 Jet A-1 fuel specification
Energy Institute Test Method IP 564
2 minute test procedure
Fully manufactured by Parker with 20 years experience in the Particle Counter Measuring market
Laser optical scanning analysis
Multi-standard ISO cleanliness reporting
On-board, rear-mounted pump enables monitoring possibilities. For example: fuel storage / vehicle tanks and fuel storage drums
Latest averaging software as standard
Downloader software

Product Benefits:
Principle of Operation: optical scanning analysis and measurement of actual particles and inference to water presence
Primary Output: ≥4µ(c), ≥6µ(c), ≥14µ(c), ≥21µ(c), ≥25µ(c), ≥30µ(c) counts per ml
Secondary Diagnostic Output: % Volume Distribution, via graphical display on handset and printout
International Codes: ISO 7-22 in accordance with ISO 4406-1999
Data Entry: 32 character two line dot matrix LCD. Full alpha numeric entry facility on keypad
Data Retrieval: memory access gives test search facility for up to 300 saved tests
Calibration: in accordance with Parker Calibration Procedure CM20-N, which complies to ISO11171-1999, Clause 6 (Ommitting Annex F)
Re-calibration: Every 12 months by a dedicated Parker Service Centre (Consult Parker) as required under strict El methods
Laboratory Sampling: Utilises on-board rear mounted pump


Technical details:


Test Time: 2 minutes
Mechanical Components: Brass, plated steel, stainless steel and aluminium
Flow Rate: 25 - 28 ml/min (dictated by CMP) 100 ml/min with additional flush button
Fluid Compatibility: Hydrocarbon Fuel, Mineral Oil. For other fluids consult Parker
Fuse: 1.25 amp fast blow fuse included for overload protection (spare supplied)
icountACM20 Technology: Flow cell, light obscuration
Repeatability / Accuracy: As per or better than ISO 11171
Coincidence: 40,000 particles per ml
Viscosity Range: 1 - 100 centistokes
Weight: 8kg
Carrying Case Weight: 5kg
Maximum Working Pressure: 420 bar
Operating Temperature: +5°C - +80°C
Memory Store: 300 test capacity
Power Requirement: 12vDC input, 6 x ’D’ Cell batteries or rechargeable battery pack

DEFSTAN 9191